Describe with suitable example; frequency measurement by Lissajous patterns on CRO. 

1 Answer

Answer :

* Initially the unknown frequency signal is connected to vertical deflection plates of CRO. * Now switch of the internal sweep generator of the horizontal deflecting system. * A standard source of frequency is applied to the horizontal deflection plates. * The frequency of standard source is adjusted now until a circular or elliptical pattern appears on CRT screen. * When such pattern is observed on a screen, it indicates that two frequencies are equal. * Practically, it is not possible to adjust the standard frequency exactly equal to unknown frequency. Hence the standard frequency is adjusted to be a multiple or submultiple of unknown frequency.

Vertical input Frequency measurement by Lissajous patterns on CRO can be explain by following example: Assume horizontal input frequency is 10 kHz for following pattern

electronics2electrical.com 748 -  .jpg

Related questions

Description : Lissajous pattern obtained on a CRO screen is a circle. Frequency of two signals are   (a) Equal (b) Unequal (c) Zero (d) Infinity 

Last Answer : a

Description : Explain how CRO is used for measurement of frequency & amplitude with suitable diagram

Last Answer : Voltage measurement/Amplitude: The most direct voltage measurement that can be made with the help of oscilloscope is the peak to peak value. The RMS value can be calculated from peak to peak value. ... is the time between two identical points of successive cycle of the waveform.

Description : The Lissajous pattern observed on screen of CRO is a straight line inclined at 45° to x axis. If X–plate input is 2 sin ωt, the Y–plate input is (1) 2 sin ωt (2) 2 sin (ωt + 45°) (3) 2 sin (ωt – 45°) (4) 22 sin ωt + 45°) 

Last Answer : The Lissajous pattern observed on screen of CRO is a straight line inclined at 45° to x axis. If X–plate input is 2 sin ωt, the Y–plate input is 2 sin ωt 

Description : The Lissajous pattern on the screen of a CRO is an ellipse with major axis in quadrant 2 and quadrant 4. Then the phase difference between two signals Can be :  (A) 270° (B) 210° (C) 180° (D) 300°

Last Answer : The Lissajous pattern on the screen of a CRO is an ellipse with major axis in quadrant 2 and quadrant 4. Then the phase difference between two signals Can be : 210°

Description : Consider the following statements with regard to Lissajous pattern on a CRO :  1. It is a stationary pattern on the CRO.  2. It is used for precise measurement of frequency of a voltage signal.  3. The ratio between frequencies ... /are correct ?  (a) 1 only (b) 2 only (c) 3 only (d) 1, 2 and 3

Last Answer : Consider the following statements with regard to Lissajous pattern on a CRO :  1. It is a stationary pattern on the CRO.  2. It is used for precise measurement of frequency of a voltage signal.  3. The ratio between ... /are correct ?  (a) 1 only (b) 2 only (c) 3 only (d) 1, 2 and 3

Last Answer : The Lissajous pattern obtained on a CRO in used to determine: Phase shift and frequency

Description : On oscilloscope screen for dynamic measurement, lissajous diagram is normally used to determine the __________ of the input signal. (A) Frequency (B) Amplitude (C) Damping (D) None of these

Last Answer : (A) Frequency

Description : Explain measurement of phase and frequency using Lissajous pattern. Write the formula for each one.

Last Answer : Phase measurement of Lissajous pattern: The phase measurement can be done by using Lissajous figures. The CRO is set to operate in the X- Y mode, then the display obtained on the screen of a ... or sub- multiple of the unknown frequency.  The Lissajous pattern appears stationary.     

Description : Define Lissajous pattern. Explain how Lissajous pattern is useful for frequency and phase measurement.

Last Answer : Lissajous Pattern : The CRO is set to operate in the X- Y mode, when two sine waves of the same frequency are applied to the CRO (One vertical and one horizontal deflection plates) then the display obtained on ... 1. Lissajous pattern for ratio 3:1    2. Lissajous pattern for ratio 3:2  

Description : What will be the phase shift for following Lissajous patterns?

Last Answer : (i) Phase shift = 0° (ii) Phase shift = 90° or 270° (iii) Phase shift = 30° or 330° (iv) Phase shift = 180°

Last Answer : A Lissajous patterns are used to measure Frequency and Phase shift.

Description : Describe the procedure for the measurement of earth resistance by using earth tester.

Last Answer : Measurement of earth resistance by using earth tester:  It is connected to earth electrode E, whose resistance is to be measured, and the other spike P and R, as shown in the ... , the deflection of the pointer will indicate directly resistance in ohm of the earth electrode under test. 

Description : Explain the measurement of voltage and frequency using CRO.

Last Answer : Voltage Measurement: CRO displays the voltage signal as a function of time on its screen. The amplitude of that voltage signal is constant, but we can vary the number of divisions that cover the voltage ... /division nv is the number of divisions that cover the signal in vertical direction.

Description : Describe with block diagram the working of digital frequency meter.

Last Answer : Digital frequency meter: Amplifier: The signal whose frequency is to be measured is first amplified and supplied to the schmitt trigger. Schmitt Trigger: The schmitt trigger convert ... count displayed on counter  t = time interval between start and stop condition of the gate 

Description : Explain the effect of power factor on wattmeter reading in two wattmeter method for power measurement.

Last Answer : Effect of P.F. on Reading of Wattmeter in Two Wattmeter Method:

Description : Draw the circuit diagram for:- i) Measurement of active power in 3-phase load circuit using two wattmeter. ii) Measurement of reactive power in 3-phase load circuit using one wattmeter.

Last Answer : (i) Measurement of active power in 3- phase load circuit using two- wattmeter: (ii) Measurement of reactive power in 3- phase load circuit using one wattmeter

Description : Compare CRO and DSO. State the formula for phase measurement using CRO with necessary diagram.

Last Answer : CRO DSO Directly reads analog voltage and displays it on screen. It reads the analog voltage and converts it into digital form before being displayed on the screen. ... algorithms available and ported on microprocessor which can operate on stored samples of input voltage.

Description : sketch time period measurement by CRO.

Last Answer : Horizontal sweep is turned ON and the display appealing on the screen is adjusted by varying different control knobs provided on the front panel of CRO, till the signal is suitably displayed ... Where m=Number of division in one complete cycle n=Setting of time base =Time/Division

Description : Describe with block diagram; the principle of operation of digital storage oscilloscope.

Last Answer : The block diagram of DSO is shown above. Sampling rate may be as high as 1,00,000 samples per sencod. For such high sampling rates, a fast conversion ADC is requried. Flash ADC ... particular signal in the memory then overwritting should be stopped.This obtained by using Hold and Save button.

Description : Describe with block diagram; the construction of single phase electronic energy meter. 

Last Answer : Construction of Single Phase Electronic Energy Meter: 1. CT reduces current to reasonable value for current scaling network. 2. Voltage & current scaling networks reduce proportionally the ... The ADC converts the square wave frequency analog output to display the energy in Watt-hour.

Description : Describe different types of errors in analog measuring instruments. 

Last Answer : Types of Errors in Analog Measuring Instruments: A) Gross error: These are due to mistakes on the part of person using the instrument. B) Systematic Error: It has three sub types as ... multiplying factor. C) Random error: These persist even after gross and systematic errors are removed.

Description : Two sinusoidal signals having the same amplitude and frequency are applied to the X and Y inputs of a CRO. The observed Lissajous figure is straight line. The phase shift between the two signals would be (1) Zero (2) 90 degrees (3) Either zero or 180 degrees (4) Either 90 degrees or 270 degrees  

Last Answer : Two sinusoidal signals having the same amplitude and frequency are applied to the X and Y inputs of a CRO. The observed Lissajous figure is straight line. The phase shift between the two signals would be (1) Zero (2) 90 degrees (3) Either zero or 180 degrees (4) Either 90 degrees or 270 degrees  

Description : A piano tuner notices a standing Lissajous (pron: Liss-a-jew) curve on his oscilloscope. The tone being projected is: w) out of harmony x) sharp y) resonating z) flat

Last Answer : ANSWER: Y -- RESONATING

Description : State significance of Lissajous figure. 

Last Answer : Significance of Lissajous figure. The characteristics patterns that appear on the screen of a cathode ray tube, when sinusoidal voltages are simultaneously applied to horizontal and vertical plates .these ... is presented as a Lissajous figure. it is used for measurement of phase and frequency.

Description : State and describe different triggering available in CRO

Last Answer : A trigger is an external signal or input signal condition that initiates one or more instrument functions. Trigger settings include digital edges, software functions, and analog levels. This tutorial ... the external trigger input. The following figure shows an entering window trigger.  

Description : Draw the basic block diagram of single trace CRO and describe the function of delay line.

Last Answer : Function of Delay line: This block is used to delay the signal for a period of time in the vertical section of CRT. The input signal is not applied directly to the vertical plates because the part ... when the delay time is not used. Therefore, the input signal is delayed by a period of time. 

Description : Draw and describe horizontal deflection system in CRO 

Last Answer : Block Diagram: Explanation: Input from vertical amplifier is given to Horizontal system. Trigger circuit: a. A trigger circuit is used to convert the incoming signal into trigger ... : Horizontal amplifier amplifies the signal & passes the signal to horizontal plates of CRT. 

Description : Explain with neat sketch the construction and working principle of megger.

Last Answer : Megger: Construction of Megger : Two coils: the current coil and pressure coil are mounted at an angle on the same spindle and form the part of the moving system. These coils are ... force produces a torque that is used to deflect the pointer of the device which gives some reading.

Description : Explain the calibration of single phase electronic energy meter using direct loading.

Last Answer : Calibration of Single-phase Electronic Energy Meter by Direct Loading Method: The following figure shows the arrangement of calibration for single phase electronic energy meter. A phase shifter is connected for ... are taken for different values of current at different power factors. 

Description : Draw block diagram of function generator and state the function of each block. 

Last Answer : Block Diagram of Function Generator: OR Working of Function Generator : This instrument can deliver sine, triangular & square waves with frequency range of 0.01 Hz to ... output voltage reaches a predetermined level on the negative slope of the integrator's output voltage. 

Description : Explain with neat sketch; the construction and working principle of Repulsion type moving iron instrument.

Last Answer : Repulsion Type Moving Iron Instrument: Construction: It consist of fixed cylinder hollow coil which carries operating current inside the coil there are two soft iron pieces. One of which is ... magnetized with same polarity a repulsive force act on them and moving iron piece gets deflected. 

Description : Draw the block diagram of tri-vector meter. State the various measurements possible from tri-vector meter.

Last Answer : Various measurements possible from tri-vector meter are: Tri-vector meter is a measuring instrument which measures the kWh, kVArh, the kVAh of a power line. Tri-vector meter is an energy meter ... of supply such as voltage, current, power factor, active load, reactive load, apparent load etc.

Description :  Draw a neat labeled block diagram of 3-phase Electronic Energy meter.

Last Answer : Labeled diagram of 3-phase electronic energy meter:

Description : Explain the error occurred due to pressure coil inductance of electrodynamometer type wattmeter. How this error is compensated?

Last Answer : For power measurement assume that the pressure coil of wattmeter is purely restitive. Hence current taken by pressure coil Ipc is in phase with voltage across the pressure coil Vpc  ... inductive reactance XL the pressure coil circuit is made purely resistive and the error is eliminated.

Description : Explain with neat sketch, the working of full wave rectifier voltmeter.

Last Answer : Full Wave Rectifier Type Voltmeter: Working of full wave rectifier type A. C. Voltmeter Referring to the circuit diagram, for rectification, a full wave bridge circuit (ABCD) is used. Input ... current passing through meter is given as, which causes full scale deflection of the meter.

Description : A permanent magnet moving coil instrument gives full scale deflection of 25mA when a voltage of 75mV is applied across it. Calculate: (i) Resistance of shunt required for full scale deflection of 150 A. (ii) Series resistance for full scale deflection of 500 V.

Last Answer : Answer :

Description : State the purpose of calibration of measuring instruments. Explain the procedure of calibration of D.C. voltmeter by using D.C. potentiometer.

Last Answer : Purpose of calibration : Calibration is the process of evaluating and adjusting the precision and accuracy of measuring equipment. Proper calibration of an instrument allows people to have a safe ... calibration curve can be drawn by plotting the percentage error against the measured voltages. 

Description : List the types of systematic errors and state the reasons due to which these errors occur.

Last Answer : Systematic Error: i) Instrumental Error: These errors are caused due to the mechanical structure of measuring instrument such as: a) Inherent shortcomings of instruments: Instrument may ... or electrostatic fields. iii) Observational Error: Parallax errors, incorrect multiplying factor.

Description : Compare analog instrument to digital instrument on the basis of accuracy, resolution, power required and portability. 

Last Answer : Comparison between Analog instruments & Digital instruments: 

Description : State the advantages of electronic energy meter. 

Last Answer : Advantages of Electronic Energy Meter: 1) Low cost. 2) High accuracy. 3) More flexibility. 4) High sensitivity 5) No frictional losses. 6) Less loading effect. 7) High frequency ... for external adjustment. 11) Low load, full load, power factor and creeping adjustments are not required.

Description : List the errors occuring in single phase electronic energy meter.

Last Answer : Errors in Single Phase Electronic Energy Meter: 1) Error due to wrong sensor opeartion. 2) Error due to temperature. 3) Error due to wrong CT opeartion. 4) Error due to wrong PT opeartion. 5 ... of scaling network. 6) Error due to wrong ADC opeartion. 7) Error due to wrong Display. 

Description : A single phase wattmeter rated for 500 V; 5A is having full scale deflection of 1000 watt. What is multiplying factor of the wattmeter?

Last Answer : Answer :

Description : State the purpose of four quadrant meter.

Last Answer : Purpose of four quadrant meter:  Four-quadrant meter is basically a bidirectional energy-meter with additional measurement facility. The purposes of four-quadrant meter are: 1) Measurement of energy/power ... and kVARh 3) Time-of-use metering: Bidirectional consumption and demand.

Description : Calculate the resistance of shunt required to make a milliammeter which gives maximum deflection for a current of 15 mA and which has a resistance of 5 Ω; read up to 10 Amp.

Last Answer : Answer :

Description : State the working principle of PMMC Analog instrument.

Last Answer : Working Principal of PMMC Analog Instrument: When current carrying conductor is placed in a magnetic field, it experiences mechanical force. In PMMC instrument, when current flows through a coil ... coil rotates and pointer attached to the coil shows deflection proportional to the current magnitude.

Description : Differentiate between analog and digital multimeter.

Last Answer : Analog multimeter Digital multimeter 1. Power supply is not required 1. Power supply is required 2. Less suffered from electric noise 2. Suffered from electric noise 3. It provides measurement ... 9. Bigger in size 9. Compact in size 10. Economical 10.Expensive

Description : “Secondary of current transformer should not be open”. Justify.

Last Answer : Secondary of a CT should not be Open Circuited: If secondary of CT is open circuited ,then current through secondary becomes zero hence the ampere-turns produced by secondary which generally opposes primary ... opeartor point of view as well. So secondary of a CT should not be open circuited. 

Description : State the use of synchroscope.

Last Answer : Use of Synchroscope: The synchroscope is used to determine the exact instant of switching required to connect the alternators in parallel or put an alternator in parallel with the infinite bus. ... idea of the faster or slower machine to make adjustments in speed of alternators for synchronising.

Description : Give ranges of low and high resistances.

Last Answer : 1) Ranges of low resistance: Less than 1 ohm.  2) Ranges of high resistance: Greater than 0.1 Mega ohm.

Description : State the meaning of energy meter constant K. State its unit.

Last Answer : Meaning of Energy Meter Constant (K): Energy Meter Constant is the no. of revolutions made by disc in meter per unit (kWh) consumption of energy. Unit of energy meter constant K is rev/kWH