A moving coil instrument of resistance 5 Ohm requires a potential difference of 75 mV to give a full scale deflection. The value of shunt resistance needed to give a full scale deflection at 30 A is

1 Answer

Answer :

A moving coil instrument of resistance 5 Ohm requires a potential difference of 75 mV to give a full scale deflection. The value of shunt resistance needed to give a full scale deflection at 30 A is 2.5 milliOhm

Related questions

Description : A potentiometer having the potential gradient of `2 mV//cm` is used to measure the difference of potential across a resistance of `10 ohm`. If a lengt

Last Answer : A potentiometer having the potential gradient of `2 mV//cm` is used to measure the difference of potential across a ... `) A. 1 B. 2 C. 5 D. 10

Description : A potentiometer having the potential gradient of `2 mV//cm` is used to measure the difference of potential across a resistance of `10 ohm`. If a lengt

Last Answer : A potentiometer having the potential gradient of `2 mV//cm` is used to measure the difference of potential across a ... `) A. 1 B. 2 C. 5 D. 10

Description : A permanent magnet moving coil ammeter has a coil resistance of 99 ohm and Full Scale Deflection (FSD) current of 0.1mA. Shunt resistance is 1 ohm. Current through the meter at 0.5 F.S.D is:

Last Answer : A permanent magnet moving coil ammeter has a coil resistance of 99 ohm and Full Scale Deflection (FSD) current of 0.1mA. Shunt resistance is 1 ohm. Current through the meter at 0.5 F.S.D is: 0.05mA

Description : A permanent magnet moving coil instrument gives full scale deflection of 25mA when a voltage of 75mV is applied across it. Calculate: (i) Resistance of shunt required for full scale deflection of 150 A. (ii) Series resistance for full scale deflection of 500 V.

Last Answer : Answer :

Description : A moving coil instrument has a resistance of 10 W and gives full scale deflection when carrying a current of 50 mA. To measure current upto 100 A, we will have to connect • 10 W resistance in series • 10 W resistance in parallel • 10/1999 W resistance in series • 10/1999 W resistance in parallel

Last Answer : A moving coil instrument has a resistance of 10 W and gives full scale deflection when carrying a current of 50 mA. To measure current upto 100 A, we will have to connect 10/1999 W resistance in parallel

Description : A coil of inductance 240 mH and resistance 75 Ohm is connected in parallel with a capacitor across a 30 V, variable frequency supply. The current drawn by the circuit is found to be minimum when the supply frequency is 1 kHz. The ... (A) 40, 400 Hz (B) 10, 100 Hz (C) 20, 50 Hz (D) 50, 20 Hz

Last Answer : A coil of inductance 240 mH and resistance 75 Ohm is connected in parallel with a capacitor across a 30 V, variable frequency supply. The current drawn by the circuit is found to be minimum ... supply frequency is 1 kHz. The approximate values of Q-factor and bandwidth are, respectively 20, 50 Hz

Description : What is the typical full-scale deflection current of a moving coil instrument?

Last Answer : 50mA

Description : The potential coil of a single phase dynamometer wattmeter has 4840 Ohm resistance. Voltage across the load is 220 V. With the potential coil connected on the load side, the meter indicates 100 W. Approximate percentage error due to wattmeter connection is (A) 11% (B) 9% (C) 1.1% (D) 3%

Last Answer : The potential coil of a single phase dynamometer wattmeter has 4840 Ohm resistance. Voltage across the load is 220 V. With the potential coil connected on the load side, the meter indicates 100 W. Approximate percentage error due to wattmeter connection is 11%

Description : What is the reason why the scale of a permanent-magnet moving coil instrument is uniform?

Last Answer : because it is spring controlled

Last Answer : A moving coil instrument is used as an ohmmeter. The indicating scale of the meter will be Hyperbolic.

Description : A galvanometer with 20 ohms coil resistence has a full-scale deflection of 10 mA. A 0.02 ohm is placed across the meter to increase its rating capacity. What is the new full scale current for the meter?

Last Answer : False.

Description : A moving coil meter has a resistance of 3 Ω and gives full scale deflection with 30 mA. To measure voltage up to 300 V, the external resistance to be connected in series with the instrument is (A) 9997 Ω (B) 10000 Ω (C) 997 Ω (D) 10003 Ω

Last Answer : A moving coil meter has a resistance of 3 Ω and gives full scale deflection with 30 mA. To measure voltage up to 300 V, the external resistance to be connected in series with the instrument is 9997 Ω

Description : A nichrome wire used as heater coil has the resistance of 1 ohm/m. For a heater of 1000 W at 200 V, the length of wire required will be 

Last Answer : A nichrome wire used as heater coil has the resistance of 1 ohm/m. For a heater of 1000 W at 200 V, the length of wire required will be 40 m.

Description : The voltage induced across a certain coil is 200 mV. A 1200 resistor is connected to the coil terminals. The induced current will be: A. 1.7 mA B. 16 mA C. 12 mA D. 120 mA

Last Answer : The voltage induced across a certain coil is 200 mV. A 1200 resistor is connected to the coil terminals. The induced current will be: 1.7 mA

Last Answer : In a dynamo-meter type of moving coil instrument a swamping resistance is provided inorder to: Provide equal time constant for moving coil and fixed coil, when used for ac measurement

Last Answer : A moving-coil permanent-magnet instrument can be used as ammeter by using a low resistance shunt.

Description : Potentiometer wire of length `1 m` is connected in series with `490 Omega` resistance and `2 V` battery. If `0.2 mV/cm` is the potential gradient, the

Last Answer : Potentiometer wire of length `1 m` is connected in series with `490 Omega` resistance and `2 V` battery. If ... 7.9Omega` C. `5.9Omega` D. `6.9Omega`

Description : Name the material used for :- (i) Moving coil (ii) Permanent magnet in PMMC instrument.

Last Answer : (i) Material used for Moving coil: Copper, Aluminum( some times)  (ii) Permanent magnets are made of special alloys such as :  Aluminum-Nickel-Cobalt (Alnicos)  Strontium-Iron  Neodymium-Iron-Boron  Samarium-Cobalt.

Description : Full scale Ammeter reading is 10 mA, Voltage across Ammeter is 100 mV. Then for 100 A measurements the power loss in the circuit is a) 1 W b) 10 W c) 100 W d) 1000 W

Last Answer : Ans: 10 W*

Description : The potential difference across the gap of a spark plug ranges between a) 10-20 mV b) 10-20 V c) 10-20 kV d) 10-20 MV

Last Answer : ans a

Description : What is the typical full-scale voltage across a moving coil voltmeter?

Last Answer : 50mV

Description : : Moving a coil in and out of magnetic field induces A. force B. potential difference C. e.m.f D. voltage

Last Answer : e.m.f

Description : A dc voltmeter has a sensitivity of 1000 Ohm/V. When it measures half full scale in 100 V range, the current through the voltmeter is

Last Answer : A dc voltmeter has a sensitivity of 1000 Ohm/V. When it measures half full scale in 100 V range, the current through the voltmeter is 1 mA.

Description : A 220 V dc machine has an armature resistance o f 1-ohm. If the full load current is 20A, the difference in the induced voltage when the machine is running as a motor, and as a generator is

Last Answer : A 220 V dc machine has an armature resistance o f 1-ohm. If the full load current is 20A, the difference in the induced voltage when the machine is running as a motor, and as a generator is 40 V.

Description : For amplitude measurements, ___ scale is calibrated in either volts per centimeter (V/cm), or millivolts per centimeter (mV/cm)

Last Answer : vertical

Description : Explain the construction and working of moving coil instrument.

Last Answer : Construction: The moving part is a coil wound on a light aluminium frame, mounted on a shaft made of stainless steel which is pivoted at either ends on jewel bearing made of sapphire. The coil is placed between ... . Where B is magnetic flux, N is number of turns and L is length of coil. 

Description : Current range extension in moving coil instruments can be achieved by placing a ___ in shunt with the instrument.

Last Answer : low resistance resistor

Description : Permanent-magnet moving coil instrument can be used in

Last Answer : dc work only

Description : Permanent Magnet Moving Coil (PMMC) Instrument

Last Answer : Permanent Magnet Moving Coil (PMMC) Instrument

Last Answer :  In a moving coil instrument, the deflecting torque is proportional to Current.

Last Answer : A moving-coil permanent-magnet instrument can be used as flux-meter by eliminating the control springs.

Description : A particle of mass m is moving along a straight path through the origin of coordinates with a velocity v. The angular momentum of the particle about the origin (a) mv (b) Depends on its distance from origin (c) Is zero (d) Changes from positive to negative as it passes through the origin

Last Answer : Ans:(c)

Description : According to the equation ‘r = (mv)⁄(Be)’, the faster moving particles A. move in smaller circle B. move straight C. move in bigger circle D. move randomly

Last Answer : move in bigger circle

Description : At rest, the interior of the typical mammalian neuronal axon potential is approximately ________ mV (a) 0 (b) –70 (c) –100 (d) –20

Last Answer : Ans: B

Description : A galvanometer has a coil of resistance `100 Omega` and gives a full-scale deflection for `30 mA` current. If it is to work as a voltmeter of `30 V` r

Last Answer : A galvanometer has a coil of resistance `100 Omega` and gives a full-scale deflection for `30 mA` current. ... Omega` C. `1000 Omega` D. `1800 Omega`

Description : A galvanometer having resistance of `50 Omega` requires a current of `100 Omega A` to given full scale deflection. How much resistance is required to

Last Answer : A galvanometer having resistance of `50 Omega` requires a current of `100 Omega A` to given full scale ... in series D. `10^(5)Omega` in parallel

Description : A PMMC instrument has FSD of 100 μA and a coil resistance of 1 kΩ. To convert the instrument into an ammeter with full scale deflection of 100 mA, the required shunt resistance is (A) 1 Ω (B) 1.001 Ω (C) 0.5 Ω (D) 10 Ω

Last Answer : A PMMC instrument has FSD of 100 μA and a coil resistance of 1 kΩ. To convert the instrument into an ammeter with full scale deflection of 100 mA, the required shunt resistance is 1.001 Ω

Description : The resistance of a coil used in a platinum-resistance thermometer at `0^(@)C` is `3.00 Omega` and at `100^(@)C` is `3.75 Omega`. Its resistance at an

Last Answer : The resistance of a coil used in a platinum-resistance thermometer at `0^(@)C` is `3.00 Omega ... as `3.15 Omega`. Calculate the unknown temperature.

Description : Moving-iron instrument has a/an scale

Last Answer : squared

Description : A multipurpose instrument for measuring current, potential difference and resistance is a) Ohmmeter b) Astatic galvanometer c) Rheostat d) AVO Meter

Last Answer : d) AVO Meter

Description : A megohmmeter is connected to each end of an individual motor winding. A low ohm reading indicates ____________. A. an open coil B. a loose coil C. good continuity D. a dirty coil

Last Answer : Answer: C

Description : The potential difference across a 4 ohm resistor is 20 volts. What is the current in the resistor?

Last Answer : ANSWER: 5 AMPS

Description : The potential difference across a 4 ohm resistor is 20 volts. Assuming that all of the energy dissipated by this resistor is in the form of heat, how many joules of heat are radiated in 10 seconds?  

Last Answer : ANSWER: 1000 JOULES

Description : The constant potential difference across a 2 ohm resistor is 20 volts. How many watts of power are dissipated by this resistor?

Last Answer : ANSWER: 200 WATTS

Description : A moving coil galvanometer is converted into an ammeter reads upto `0.03 A` by connecting a shunt of resistance `4r` across it and ammeter reads up `0

Last Answer : A moving coil galvanometer is converted into an ammeter reads upto `0.03 A` by connecting a shunt of resistance `4r` ... `0.02A` C. `0.03A` D. `0.04A`

Description : To send 10% of the main current through a moving coil galvanometer of resistance `99omega`, the shunt required is –

Last Answer : To send 10% of the main current through a moving coil galvanometer of resistance `99omega`, the shunt ... `11Omega` C. `10Omega` D. `9.9Omega`

Description : To limit the current flow through a DC voltmeter to as low a value as possible, the moving coil circuit is provided with a/an _____________. A. high series resistance B. high parallel resistance C. series inductor D. external shunt

Last Answer : Answer: A

Description : A small swamping resistance is connected in series with operating coil of a moving coil ammter in order to compensate for the effects of ____

Last Answer : temperature variation

Last Answer : The resistance in the circuit of the moving coil of a dynamometer wattmeter should be high.

Description : The true power indicated by a wattmeter depends on the current flow through the load, the magnitude of the potential across the load, and the _____________. A. power factor of the load B. angle of coil displacement C. inertia of the movable coil D. high resistance from the load

Last Answer : Answer: A