Explain the construction and working of moving coil instrument.

1 Answer

Answer :

Construction: The moving part is a coil wound on a light aluminium frame, mounted on a shaft made of stainless steel which is pivoted at either ends on jewel bearing made of sapphire. The coil is placed between poles of a permanent magnet. Moving system is made light as far as possible to have high torque and weight ratio. There are two phosphor bronze springs of very less resistance. This is used as terminals for the current to pass through the coil and out of coil. It is also serves the purpose of control force. Aluminium coil former acts as a damping device by eddy current damping. There is a knife edge pointer and a counter weight to avoid its sagging. 

Working: When ever a current carrying conductor is placed in a magnetic field a force is experiences by the conductor. Moving coil meters work on this principle.

Force = BINL

As the BNL is constant, force (F) is directly proportional to current (I).

That is FαI. Where B is magnetic flux, N is number of turns and L is length of coil. 


Related questions

Description : Explain with neat sketch; the construction and working principle of Repulsion type moving iron instrument.

Last Answer : Repulsion Type Moving Iron Instrument: Construction: It consist of fixed cylinder hollow coil which carries operating current inside the coil there are two soft iron pieces. One of which is ... magnetized with same polarity a repulsive force act on them and moving iron piece gets deflected. 

Description : Current range extension in moving coil instruments can be achieved by placing a ___ in shunt with the instrument.

Last Answer : low resistance resistor

Description : Permanent-magnet moving coil instrument can be used in

Last Answer : dc work only

Description : What is the reason why the scale of a permanent-magnet moving coil instrument is uniform?

Last Answer : because it is spring controlled

Description : What is the typical full-scale deflection current of a moving coil instrument?

Last Answer : 50mA

Description : Permanent Magnet Moving Coil (PMMC) Instrument

Last Answer : Permanent Magnet Moving Coil (PMMC) Instrument

Description : A permanent magnet moving coil instrument gives full scale deflection of 25mA when a voltage of 75mV is applied across it. Calculate: (i) Resistance of shunt required for full scale deflection of 150 A. (ii) Series resistance for full scale deflection of 500 V.

Last Answer : Answer :

Description : Name the material used for :- (i) Moving coil (ii) Permanent magnet in PMMC instrument.

Last Answer : (i) Material used for Moving coil: Copper, Aluminum( some times)  (ii) Permanent magnets are made of special alloys such as :  Aluminum-Nickel-Cobalt (Alnicos)  Strontium-Iron  Neodymium-Iron-Boron  Samarium-Cobalt.

Description : A moving coil instrument has a resistance of 10 W and gives full scale deflection when carrying a current of 50 mA. To measure current upto 100 A, we will have to connect • 10 W resistance in series • 10 W resistance in parallel • 10/1999 W resistance in series • 10/1999 W resistance in parallel

Last Answer : A moving coil instrument has a resistance of 10 W and gives full scale deflection when carrying a current of 50 mA. To measure current upto 100 A, we will have to connect 10/1999 W resistance in parallel

Description : A moving coil instrument of resistance 5 Ohm requires a potential difference of 75 mV to give a full scale deflection. The value of shunt resistance needed to give a full scale deflection at 30 A is

Last Answer : A moving coil instrument of resistance 5 Ohm requires a potential difference of 75 mV to give a full scale deflection. The value of shunt resistance needed to give a full scale deflection at 30 A is 2.5 milliOhm

Last Answer :  In a moving coil instrument, the deflecting torque is proportional to Current.

Last Answer : A moving coil instrument is used as an ohmmeter. The indicating scale of the meter will be Hyperbolic.

Last Answer : In a dynamo-meter type of moving coil instrument a swamping resistance is provided inorder to: Provide equal time constant for moving coil and fixed coil, when used for ac measurement

Last Answer : A moving-coil permanent-magnet instrument can be used as flux-meter by eliminating the control springs.

Last Answer : A moving-coil permanent-magnet instrument can be used as ammeter by using a low resistance shunt.

Description : Briefly explain the construction and working of an induction coil.

Last Answer : Construction: An induction coil consists of a primary P having few turns of thick insulated copper wire wound over a soft iron core connected to a battery. It also has a secondary coil S of larger number ... repeats. During the make break of the circuit, large e.m.f is induced in the secondary coil.

Description : Explain construction & working of attraction type M.I. instrument with neat diagram.

Last Answer : 1. Moving element: a small piece of soft iron in the form of a vane or rod.  2. In Attracted type a cylindrical coil or solenoid which is kept fixed. An oval-shaped soft-iron piece are ...   9. When = the pointer will be stable & we can get correct reading of the electrical quantity

Description : To limit the current flow through a DC voltmeter to as low a value as possible, the moving coil circuit is provided with a/an _____________. A. high series resistance B. high parallel resistance C. series inductor D. external shunt

Last Answer : Answer: A

Description : Which of the following electric meter movements uses a stationary permanent magnet and movable coil? A. D'Arsonval B. Electrodynamometer C. Moving iron-vane D. Inclined coil iron-vane

Last Answer : Answer: A

Description : A PMMC instrument has FSD of 100 μA and a coil resistance of 1 kΩ. To convert the instrument into an ammeter with full scale deflection of 100 mA, the required shunt resistance is (A) 1 Ω (B) 1.001 Ω (C) 0.5 Ω (D) 10 Ω

Last Answer : A PMMC instrument has FSD of 100 μA and a coil resistance of 1 kΩ. To convert the instrument into an ammeter with full scale deflection of 100 mA, the required shunt resistance is 1.001 Ω

Description : The coil of a moving coil galvanometer has an effective area of 4xx10^(-2)m^(2)`. It is suspended in a magnetic field of `5xx10^(-2)` Wb `m^(-2)`. If

Last Answer : The coil of a moving coil galvanometer has an effective area of 4xx10^(-2)m^(2)`. It is suspended in a ... D. current sensitivity is `40 deg` `A^(-1)`

Description : In order to increase the sensitivity of a moving coil galvanometer, one should decrease

Last Answer : In order to increase the sensitivity of a moving coil galvanometer, one should decrease A. the strength of ... in its coil D. the area of its coil

Description : In a moving coil galvanometer, the deflection of the coil q is related to the electrical current i by the relation

Last Answer : In a moving coil galvanometer, the deflection of the coil q is related to the electrical current i by the ... theta^(2)` D. `I prop sqrt(theta)`

Description : A moving coil galvanometer has 150 equal divisions. Its current sensitivity is 10-divisions per milliampere and voltage sensitivity is 2 divisions per

Last Answer : A moving coil galvanometer has 150 equal divisions. Its current sensitivity is 10-divisions per milliampere and voltage ... `10^(3)` D. `10^(5)`

Description : A moving coil galvanometer is converted into an ammeter reads upto `0.03 A` by connecting a shunt of resistance `4r` across it and ammeter reads up `0

Last Answer : A moving coil galvanometer is converted into an ammeter reads upto `0.03 A` by connecting a shunt of resistance `4r` ... `0.02A` C. `0.03A` D. `0.04A`

Description : To send 10% of the main current through a moving coil galvanometer of resistance `99omega`, the shunt required is –

Last Answer : To send 10% of the main current through a moving coil galvanometer of resistance `99omega`, the shunt ... `11Omega` C. `10Omega` D. `9.9Omega`

Description : A device which converts chemical energy into electrical energy is called -----? A. motor B. generator C. moving-coil meter D. battery (Answer)

Last Answer : D. battery (Answer)

Description : : Moving a coil in and out of magnetic field induces A. force B. potential difference C. e.m.f D. voltage

Last Answer : e.m.f

Description : In generators, the rate of change of flux linkage is maximum when the coil is moving through the A. vertical position B. horizontal position C. diagonal position D. at an angle of 45

Last Answer : horizontal position

Description : Moving coil in dynamometer wattmeter connected a) In series with fixed coil b) Across supply c) In series with load d) Across load

Last Answer : Ans: Across supply

Description : What is the mathematical formula for computing the output voltage from a moving coil in a magnetic field?

Last Answer : 8. e = Emax sine θ.

Description : A small swamping resistance is connected in series with operating coil of a moving coil ammter in order to compensate for the effects of ____

Last Answer : temperature variation

Description : In a moving coil ammeter, a ___ is connected in series with the coil to compensate for temperature variations

Last Answer : swamping resistors

Description : Meters with moving coil are normally used for measuring ____

Last Answer : dc only

Description : What is the typical full-scale voltage across a moving coil voltmeter?

Last Answer : 50mV

Description : The induced current can be increased by: a. Using strong magnetic field b. moving loop faster c. Replacing loop by coil d. all of above

Last Answer : d. all of above

Description : The pole faces of U-shaped magnet in a moving coil galvanometer are made concave to a) decrease the strength of magnetic field b) create a non-uniform magnetic field c) make the magnetic field radial d) All of these

Last Answer : c) make the magnetic field radial

Description : State the material used for moving coil and former for PMMC instruments. 

Last Answer : Material Used for Moving Coil and Former for PMMC Instruments:  Material used for moving coil: Copper  Material used for Former: Aluminium

Description : A permanent magnet moving coil ammeter has a coil resistance of 99 ohm and Full Scale Deflection (FSD) current of 0.1mA. Shunt resistance is 1 ohm. Current through the meter at 0.5 F.S.D is:

Last Answer : A permanent magnet moving coil ammeter has a coil resistance of 99 ohm and Full Scale Deflection (FSD) current of 0.1mA. Shunt resistance is 1 ohm. Current through the meter at 0.5 F.S.D is: 0.05mA

Description : Permanent magnet moving coil instruments are used for

Last Answer : PMME are mostly used for measuring voltage and current.

Description : A moving coil of meter has 100 turns, and a length and depth of 10mm and 20mm respectively. It is positioned in a uniform radial flux density of 200 mT. The coil carrier current capacity is 50 mA. The torque on the coil is

Last Answer : A moving coil of meter has 100 turns, and a length and depth of 10mm and 20mm respectively. It is positioned in a uniform radial flux density of 200 mT. The coil carrier current capacity is 50 mA. The torque on the coil is 200µNm

Description : A moving coil meter has a resistance of 3 Ω and gives full scale deflection with 30 mA. To measure voltage up to 300 V, the external resistance to be connected in series with the instrument is (A) 9997 Ω (B) 10000 Ω (C) 997 Ω (D) 10003 Ω

Last Answer : A moving coil meter has a resistance of 3 Ω and gives full scale deflection with 30 mA. To measure voltage up to 300 V, the external resistance to be connected in series with the instrument is 9997 Ω

Description :  Moving coil in dynamometer wattmeter connected : - a) in series with fixed coil b) across supply c) in series with load d) across load

Last Answer : Moving coil in dynamometer wattmeter connected : - across supply

Description :  The main sources of Errors in Permanent Magnet moving Coil (PMMC) are due to : - a) weaking of the permanent magnet due to aging at temperature effects b) weaking of spring due to aging and temperature effects c) change of resistance of the moving coil with temperature d) all the above

Last Answer :  The main sources of Errors in Permanent Magnet moving Coil (PMMC) are due to : - a) weaking of the permanent magnet due to aging at temperature effects b) weaking of spring due to aging and temperature effects c) change of resistance of the moving coil with temperature d) all the above

Last Answer : Dynamometer type moving coil instruments are provided with pneumatic damping

Last Answer : The coil of a moving coil meter is would on An aluminium frame.

Last Answer : In moving coil meters, damping is provided by eddy current disk.

Last Answer : In Weston synchronoscope the moving coil is connected across bus-bars.

Last Answer : The resistance in the circuit of the moving coil of a dynamometer wattmeter should be high.

Description : 1. Which type of coal is difficult to light in the open air? 2. Leukaemia or blood cancer is characterised by abnormal increase of which cells? 3. Howmany subjects are in the Union list? ... ship was sailing? 20. Which gave an impetus to the growth of militant nationalism on a widespread scale?

Last Answer : Answer : 1. Peat 2. White blood cells 3. 97 subjects 4. Seismograph 5. Akbar 6. Loss of electrons 7. Bangladesh 8. Under valuation of the assets. 9. Volcanoes 10. Guru Gobind Singh 11. ... 15. Wandiwash 16. Western Australia 17. Kathak 18. Pitcher plant 19. Mediterranean Sea 20. Partition of Bengal