What is pressure calibration? State stepwise procedure to test the accuracy of pressure gauge with dead weight tester.

2 Answers

Answer :

Pressure Calibration: It is the process of adjusting the instrument’s output signal to match a known range of pressures. Calibration of a given pressure gauge can be done with the help of dead weight pressure gauge tester. In this method, the output of the given faulty or uncalibrated pressure gauge is compared with pre calibrated dead weight pressure and corrected. 

Diagram For Dead weight Pressure gauge tester: 

image

Stepwise procedure: 

• The handle of the dead weight tester is fully drawn out and the oil is allowed to enter in the cylinder (i.e. gauge and piston) 

• A known accurate weight is placed on the platform. The area of the piston is also known; hence we can calculate the pressure. 

• Now the handle is turned in clockwise direction so that the pressure will buildup on the gauge side as well as platform side. 

• The pressure is increased by rotating the handle clockwise until enough pressure is developed inside the cylinder to lift the platform with weights placed on it and it floats freely. 

• The procedure is repeated for different weights (increase weights in steps). In the same way most of the pressure gauges are calibrated against dead weight testers. 

Answer :

Explanation- The handle is fully drawn out and the oil is allowed to enter in the cylinder (i.e. gauge and piston). A known accurate weight is placed on the platform. The area of piston is also known; hence we can calculate the pressure. Now the handle is turn to press in clockwise direction so that the pressure will buildup on the gauge side as well as platform side. Increase the pressure by rotating the handle clockwise until enough pressure is developed inside the cylinder and lifts the platform with weights placed on it and it floats freely within the limit stops. Repeat the same procedure for different weights. In the same way most of the pressure gauge are calibrated against dead weight testers. An error in dead weight tester is less than 0.1% in order to reduce the friction between the piston and cylinder, the piston is gradually rotated while a reading being taken

Related questions

Description : Explain the process of calibration of pressure gauge by Dead Weight Tester.

Last Answer : Deadweight Testers (DWT) are the primary standard for pressure measurement. A dead weight tester is an instrument that calibrates pressure by determining the weight of force divided by the area the force ... at each gauge reading and ensure that it is within the acceptable accuracy limits.

Description : Describe construction and working of Dead Weight Tester used in pressure calibration.

Last Answer : Dead weight tester:  Construction: It consists of a very accurately machined, bored and finished piston which is inserted into a close-fitting cylinder. The cross sectional areas of ... within the cylinder when the fluid gauge pressure equals the dead weight divided by the piston area.  

Description : Working principle of dead weight pressure gauge tester is based on (A) Pascal's law (B) Dalton's law of partial pressure (C) Newton's law of viscosity (D) Avogadro's hypothesis

Last Answer : Answer: Option A

Description : State how pressure measurement can be done using Dead weight tester.

Last Answer : The handle is fully drawn out and the oil is allowed to enter in the cylinder (i.e. gauge and piston) The weight to be measured is placed on the platform. Now the handle is turned ... of the area of the piston and the pressure indicated on the gauge is equal to the weight under measurement.

Description : State the Thevenin’s theorem. Also write stepwise procedure for applying Thevenin’s theorem to simple circuits.

Last Answer : Thevenin's Theorem: Any two terminal circuit having number of linear impedance and sources (voltage, current, dependent, independent) can be represented by a simple equivalent circuit consisting of a single voltage ... 's equivalent circuit: Step V: Determination of Load current:

Description : The calibration of an HPLC detector can be evaluated by measurement of? A. Retention time B. Temperature C. Flow rate accuracy D. Intensity of the peak

Last Answer : D. Intensity of the peak

Description : Describe stepwise construction procedure for water bound macadam roads.

Last Answer : The construction procedure of W.B.M. road is discussed below: 1. Preparation of subgrade : The subgrade is prepared to the required grade and camber. 2. Preparation of the base course : After preparing ... by rolling or tamping. 5. Open to traffic : After drying, the road is opened to traffic.

Description : Describe stepwise construction procedure of cement concrete road by continuous bay method.

Last Answer : Continuous Construction method: In this method of construction, all the bays or slabs (ABCD) of strip are constructed continuously without any break as shown in figure. In this method, construction joints are, ... be diverted on the other half of the road. This method is also very rapid.

Description : Describe stepwise procedure to prepare the layout of a small building using total station.

Last Answer : Layout of small building by using total station: 1. On the plan supplied by an architect, number the column serially from left to right and top to bottom starting from top left ... of setting out by comparing the diagonal distance between the extreme column centres to their calculated values.

Description : Describe the stepwise procedure for painting of used vehicle.

Last Answer : Painting procedure for used vehicle: Remove dent using denting tools and dent removing procedure. Preparing the Surface: Begin by sanding the car's surface with a dual action sander and 120 grit sandpaper ... the car. Use circular and back and forth motions until the entire car has been polished.

Description : Explain the stepwise procedure of estimating volume of reservoir from any contour map.

Last Answer : Reservoirs are made for water supply and for power or irrigation projects. A contour map is very useful to study the possible location of a dam and the volume of water to be confined. All ... reservoir can be estimated by the application of either the trapezoidal or the prismoidal formula.

Description : Explain with neat sketch stepwise procedure to determine bulk density by sand replacement method.

Last Answer : i) A flat area, approximately 450sq.mm of the soil to be tested should be exposed and trimmed down to a level surface, preferably with the aid of the scraper tool. ii) The metal tray with a ... gram. These measurements should be repeated at least thrice and the mean weight (W3) taken.    

Description : Define specific gravity of soil. Explain the stepwise procedure of determination of specific gravity of soil by pycnometer. 

Last Answer : Specific Gravity: It is ratio of unit weight in air of a given volume of soil solids at a specific temperature to unit weight in air of an equal volume of distilled water at same temperature. ... Repeat all above steps two more times to calculate average specific gravity of given soil sample.

Description : Write stepwise procedure for converting raw material into finished product.

Last Answer : Planning of process from raw material to finished product:- Steps involved in process planning: 1. Analysis of Product and its specification: - During analysis, following parameter should be considered, ... fixing the rates of payment. 11. Prepare the route sheet and operation sheet.

Description : Explain stepwise the “Detailed energy audit” procedure.

Last Answer : A) Start up meeting: For this programme, we proceed with this meeting. If then continue us until implementation of energy saving measures. B) Analysis of energy used: Identify where ... are considered, the analysis of energy loss and measurement of insulation level is also considered.

Description : Describe procedure for calibration of ammeter with diagram.

Last Answer : Procedure for Calibration of Ammeter:  In this method, DC potentiometer is used for measurement of voltage across a standard low resistance.  Connect the circuit as shown in the figure, ... S = resistance of potentiometer  And the currents are compared for finding out error in ammeter.

Description : If a gauge is installed perpendicular to the slope, its measurement is reduced by multiplying (A) Sine of the angle of inclination with vertical (B) Cosine of the angle of inclination with vertical (C) Tangent of the angle of inclination with vertical (D) Calibration coefficient of the gauge

Last Answer : Answer: Option B

Description : State the purpose of calibration of measuring instruments. Explain the procedure of calibration of D.C. voltmeter by using D.C. potentiometer.

Last Answer : Purpose of calibration : Calibration is the process of evaluating and adjusting the precision and accuracy of measuring equipment. Proper calibration of an instrument allows people to have a safe ... calibration curve can be drawn by plotting the percentage error against the measured voltages. 

Description : Explain the working and principle of dead weight tester.

Last Answer : Dead weight tester:  Principle: It works on the principle that the downward force of the weight on the top of the piston is balanced by the pressure exerted by the fluid beneath the piston  ... within the cylinder when the fluid gauge pressure equals the dead weight divided by the piston area.

Description : Draw and explain the construction and working of Dead Weight Tester.

Last Answer : Construction:- Deadweight Tester (DWT) is used for calibration of pressure gauges . A dead weight tester is an instrument that calibrates pressure by determining the weight of force divided by ... the error at each gauge reading and ensure that it is within the acceptable accuracy limits.

Description : Draw labeled dead weight tester.

Last Answer : dead weight tester

Description : need of calibration and explain the procedure to calibrate the instrument.

Last Answer : Need of calibration: Every measuring device degrades over time due to normal wear and tear. These changes can be caused by various reasons such as electric or mechanical shock or a hazardous manufacturing ... different magnitude are used to calculate the resistance based on ohm's law.

Description : Describe the procedure for the measurement of earth resistance by using earth tester.

Last Answer : Measurement of earth resistance by using earth tester:  It is connected to earth electrode E, whose resistance is to be measured, and the other spike P and R, as shown in the ... , the deflection of the pointer will indicate directly resistance in ohm of the earth electrode under test. 

Description : State the principle of calibration.

Last Answer : The process of deriving the value of a quantity by comparing that quantity with a standard quantity is called as calibration. Calibration of instrument is done to obtain correct unknown value of each scale reading on measuring instrument 

Description : State two purposes of calibration of measuring instruments.

Last Answer : Purposes of calibration of measuring instruments: - The purpose of calibration is to ensure that readings of an instrument are consistent with other instruments. It is also important in determining the ... will need to be calibrated periodically to make sure it can function properly and safely. 

Description : What calibration label is used when actual measurement values must be known to use the test equipment?

Last Answer : CALIBRATED—REFER TO REPORT.

Description : What is not included in prevention costs? a) quality planning b) formal technical reviews c) test equipment d) equipment calibration and maintenance

Last Answer :  Ans: d

Description : What is a scale calibration weight?

Last Answer : A scale calibration weight allows a person to check that a set of scales are reading the correct figures. This works by placing a calibration weight of which its exact weight is knows and checking that the scale reads the same weight. If not, then the scales will need to be calibrated.

Description : State working principle of earth tester.

Last Answer : Earth tester:-The resistance of the earth is measured with the help of earth tester. It consists of a hand cranked type generator, the magnet poles, crossed coils and flexible spirals. It consists ... deflection of the pointer will indicate directly resistance in ohm of the earth under test.

Description : During oxidation energy released in a series of slow stepwise reaction controlled by (A)____ and is trapped as (B)____ in the form of (C)____

Last Answer : During oxidation energy released in a series of slow stepwise reaction controlled by (A)____ and is trapped as (B)____ in the form of (C)____

Description : During oxidation energy released in a series of slow stepwise reaction controlled by (A)____ and is trapped as (B)____ in the form of (C)____

Last Answer : During oxidation energy released in a series of slow stepwise reaction controlled by (A)____ and is trapped as (B)____ in the form of (C)____

Description : In step growth polymerisation, condensation occurs in a stepwise manner with or without the elimination of smaller molecules. An example of step growth polymerisation product is (A) Terylene (B) Polybutadiene (C) PVC (D) Polypropylene

Last Answer : (A) Terylene

Description : The overall goal of glycolysis, Krebs’ cycle and the electron transport system is the formation of (a) ATP in one large oxidation reaction (b) sugars (c) nucleic acids (d) ATP in small stepwise units

Last Answer : d) ATP in small stepwise units.

Description : Discuss in brief the stepwise process of percolation used in preparation of tincture.

Last Answer : Stages involved in preparation of tincture by percolation:  a. Imbibition: Drug is kept is moistened with sufficient quantity of menstruum. Allow to stand for 4 hr. Significance: o It allow ... extraction. After collecting 3/4th volume of product then marc is pressed. Mix the liquids.

Description : You are using a mechanical voltage tester to check the fuses in a three-phase circuit supplying an electric motor. For which of the following reasons must the motor be stopped during the test? A. ... circuit would be indicated by the voltage tester with the circuit energized. D. All of the above.

Last Answer : Answer: B

Description : To conduct an in-circuit test of a transistor, you should use a/an ____________. A. voltmeter or transistor tester B. impedance meter C. ohmmeter or transistor tester D. wattmeter

Last Answer : Answer: A

Description : Which of the following actions must be carried out before a voltage tester can be used to test the three line fuses to a three-phase motor? A. The fuses must be removed from the circuit. B. The ... disconnected and tagged. D. Nothing need be done as long as the motor is running under a light load.

Last Answer : Answer: B

Description : During the testing of a module tester X' finds a bug and assigned it to developer. But developer rejects the same, saying that it's not a bug. What X' should do? A. Report ... same bug to another developer D. Send to the detailed information of the bug encountered and check the reproducibility

Last Answer : D. Send to the detailed information of the bug encountered and check the reproducibility

Description : During the testing of a module tester X' finds a bug and assigned it to developer. But developer rejects the same, saying that it's not a bug. What X' should do? A. Report ... same bug to another developer D. Send to the detailed information of the bug encountered and check the reproducibility

Last Answer : D. Send to the detailed information of the bug encountered and check the reproducibility

Description : A test manager wants to use the resources available for the automated testing of a web application. The best choice is A. Test automater, web specialist, DBA, test lead B. Tester, test automater, web ... C. Tester, test lead, test automater, DBA D. Tester, web specialist, test lead, test automater

Last Answer : B. Tester, test automater, web specialist, DBA

Description : The following defines the statement of what the tester is expected to accomplish or validate during testing activity A. Test scope B. Test objective C. Test environment D. None of the above

Last Answer : B. Test objective

Description : Which of the following is a myth in testing? a) Tester can find bugs b) Any user can test software c) Missed defects are not due to testers d) Complete testing is not possible

Last Answer : Ans : b

Description : Which of the following are facts about a top-down software testing approach? I. Top-down testing typically requires the tester to build method stubs. II. Top-down testing typically requires the tester to build test drivers. (1) only I (2) Only II (3) Both I and II (4) Neither I nor II

Last Answer : (1) only I

Description : Define absolute pressure and gauge pressure. State the different units for pressure measurement.

Last Answer : Definition: Absolute pressure: Absolute pressure is defined as actual total pressure including atmospheric pressure actin On a surface. PAbsolute= PAtmospheric + PGauge Gauge pressure: Gauge pressure is defined ... mm of mercury column (blood pressure) inches of water Barmillibar (weather) Torr

Description : calibration and the need of calibration for measuring instruments.

Last Answer : Calibration - It is a process of estimating the value of a quantity by comparing that quantity with a standard quantity.    Need of calibration: - Calibration defines the accuracy and ... aim of calibration is to minimize any measurement uncertainty by ensuring the accuracy of test equipment.

Description : Explain need of calibration of any instrument.

Last Answer : Calibration is the comparison between measurements: between a standard known value and the unknown value.  There are three main reasons for having instruments calibrated: 1. To ensure readings from an instrument ... . 3. To establish the reliability of the instrument i.e. that it can be trusted.

Description : Define calibration. Explain process of calibration in detail. 

Last Answer : The process of deriving the value of a quantity by comparing that quantity with a standard quantity is called as calibration. Calibration of instrument is done to obtain correct unknown value of each ... instrument reading. To establish the reliability of the instrument i.e. it can be trusted. 

Description : Explain principles of calibration.

Last Answer : The process of deriving the value of a quantity by comparing that quantity with a standard quantity is called as calibration. Calibration of instrument is done to obtain correct unknown value of each scale reading on measuring instrument

Description : Draw the calibration chain.

Last Answer : Field instruments are calibrated using master instruments. Master instruments are instruments with more accuracy and greater repeatability. Master instruments are calibrated periodically at external laboratories. ... below represents the block diagram of calibration chain and traceability. 

Description : Define: (i) Resolution, (ii) Calibration, with reference to electrical measuring systems.

Last Answer : Resolution The smallest increment in input (the quantity being measured) which can be detected with certainty by an instrument is called its resolution or discrimination.  OR It is ... the measuring instrument with standard instrument to find out the error in the instrument under test.